未标题-1
未标题-1

JC-Series High and Low Temperature (Humidity) Test Chamber

Model Description

JC-Series High and Low Temperature (Humidity) Test Chamber

  • JC-Temperature (Humidity) Series
  • 330 – Chamber volume (unit: L)
  • C-minimum temperature -70°C/no C-minimum temperature -40°C

 

Product Feature

The whole system is equipped with 10-inch color capacitive touch screen as standard, which provides better user experience and improves the convenience of user touch operation.

 

Using independent intelligent control logic system, can more accurately control the temperature and humidity curve, with a wider range of temperature and humidity capabilities (see humidity MAP chart), to meet the user’s diverse testing needs.

 

The inner chamber is made of SUS304 mirror stainless steel, which has excellent corrosion resistance, easy to clean surface, and more convenient maintenance.

 

Distinctive air circulation system and software control system can minimize the water consumption of the equipment in the mode of humid and hot working conditions (without recycling water). Under the same operating conditions, the water consumption per unit of time demonstrates excellent water-saving performance.

 

Advanced energy-saving design, the use of intelligent control system + special air circulation system can automatically adjust the operating parameters of the equipment to achieve effective use of energy.

JC-Series High and Low Temperature (Humidity) Test Chamber
Model   JC330(C)   JC610(C) JC1000(C) JC1500(C) JC2000(C)
Nominal inside volume(L) 335 605 1000 1530 2000
Internal dimension(mm) D 760 795 960 1470 1670
W 585 800 1100 1100 1100
H 750 950 950 950 1100
External dimension(mm) D 1820 1850 2000 2525 2700
W 880 1100 1450 1450 1430
H 1950 2230 2230 2230 2400
Performance Parameters Minimum temperature ℃ End with C: minimum temperature -70℃; without C: minimum temperature -40℃
Maximum temperature ℃ +180℃(Maximum temperature can be up to 220℃ according to user’s actual demand)
Temperature fluctuation ℃ ±0.1℃~±0.5℃
Temperature uniformity ℃ ≦2℃(≤150℃)
Temperature deviation ℃ ≦±2℃
Temperature rise rate ℃/min 4.0 6.0 4.0 4.0 3.0
Cooling rate ℃/min 3.5 2.6 2.8 2.5 2.0
Humidity range %R.H. 10%~+98% R.H. (humidity operating range +10°C~+95°C)
Dew point temperature range ℃ 0~94℃
Humidity deviation %R.H. -3%R.H.~+2%R.H.
Humidity fluctuation %R.H. ≤ ± 3 %R.H.
Test method The above indexes are measured at no load, ambient temperature 20±2℃ (according to GB/T 5170.2-2017)
Rate assessment range:Tmin-Tmax, according to IEC60068-3-5, measured at the air outlet, empty chamber indicator.
Temperature and humidity performance test according to standard IEC60068-3-5, the sensor is placed at the air outlet of the air handling unit.
Meets standards ①Low Temperature Test: IEC 60068-2-1; IEC 60721-4; MIL-STD-810 G, Meth.502.5; GB/T2423.1 TestA
②High temperature test: IEC 60068-2-2; IEC 60721-4; MIL-STD-202 G, Meth.108A; MIL-STD-810 G, Meth.501.5; MIL-STD-883 J, Meth.1008.2; GB/T 2423.2 Test B
③Temperature and humidity tests: IEC 60068-2-67; IEC 60068-2-78; IEC 60721-4; ISO 16750-4, Damp heat steady; GB/T 2423.3 Test Cab; GB/T 2423.4 Test Db -Test cycle-Variant 1 and Variant 2; GB/T 2423.50 Test Cy; lEC 60068-2-30 Test Db -Test cycle -Variant 1 and Variant 2
Thermal compensation capacity 1500-3000 W
Cooling method 400W (temperature 25°C to 90°C, humidity not exceeding 90%)

Contact

  • SICHUAN JUPITER TECHNOLOGY CO.,LTD
  • +86 13601818584ly@jupiter-zyp.com
  • Room 207 No. 3699 Gonghexin Road Shanghai China, Shanghai, China
  • Leave message

    Request Inquery