Introduction
Such a series of impact test chamber adopts a structure of cold and thermal chambers and utilizes a bucket rapidly transforming specimens to the test area, so as to conduct cold and thermal shock tests.
This impact mode reduces thermal load during cold and thermal impact and shortens temperature recovery time to the greatest extent.
Feature
Temperature Shock Chamber(Two Chamber) | |||||||||
Model | JTS50M-B/S/E | JTS100M-B/S/E | JTS150M-B/S/E | JTS200M-B/S/E | JTS320M-B/S/E | JTS500M-B/S/E | JTS1000M-B/S/E | ||
Nominal inside volume(L) | 50L | 100L | 150L | 200L | 320L | 500L | 1000L | ||
Performance Parameter | Heating chamber | Temperature exposure range | +50℃~+150℃ | ||||||
Upper limit of preheating temperature | 200℃ | ||||||||
Heating time | RT~+200℃, B type: ≤ 50min; S type: ≤ 40min; E type: ≤ 30min | ||||||||
Cooling chamber | Temperature exposure range | B type: 0~-40℃; S type: 0~-55℃; E type: 0~-65℃ | |||||||
Lower limit of pre cooling temperature | -75℃ | ||||||||
Cooling time | RT~+70℃, B type: ≤ 60min; S type: ≤ 50min; E type: ≤ 40min | ||||||||
Hand basket | Temperature shock range | B type:-40℃~+150℃; S type: -55℃~+150℃; E type: -65℃~+150℃ | |||||||
Temperature fluctuation | ≤±0.5 | ||||||||
Temperature uniformity | ≤2℃ | ||||||||
Temperature deviation | ≤±2℃ | ||||||||
Conversion time | ≤10s | ||||||||
Temperature recovery time | 5 min | ||||||||
Load (steel) | 3kg | 5kg | 7kg | 10kg | 15kg | 20kg | 40kg | ||
Optional: High and low temperature test function | Temperature range | -65℃~+150℃ | |||||||
Temperature rise and fall rate | B type nearly 2℃/min; S type nearly 4℃/min; E nearly 6℃/min | ||||||||
Remarks | Above indicators are measured in no load at ambient temperature of 20±2°C(According to GB/T 5170.2-2017) | ||||||||
Inner dimensions | D | 350 | 400 | 500 | 600 | 650 | 800 | 1200 | |
W | 350 | 500 | 600 | 600 | 600 | 700 | 800 | ||
H | 400 | 500 | 500 | 550 | 550 | 900 | 1000 | ||
External dimensions | D | 2000 | 2250 | 2350 | 2550 | 1950 | 2200 | 2650 | |
W | 1400 | 1570 | 1670 | 1670 | 2700 | 3000 | 3000 | ||
H | 1950 | 2150 | 2150 | 2250 | 2300 | 2400 | 2600 | ||
Caster | 4, 6 for over 500L (caster dimension is not included in above overall dimension,caster rises by 60-120mm) | ||||||||
Test hole | OneΦ50 | OneΦ100 | |||||||
Stainless steel shelf (2 pcs in standard configuration) | Bearing 15kg/pc | Bearing 30kg/pc | Bearing 50kg/pc | ||||||
Using environment temperature | 0~+40℃ | ||||||||
Shell | Sprayed steel plate | ||||||||
Interior | SUS304 stainless steel | ||||||||
Cooling mode | water-cooling, (water temperature 7~28℃, hydraulic pressure 0.2~0.4Mpa,To ensure cooling performance) | ||||||||
Controller | 7′ Color touch screen Controller | ||||||||
Communication function | RS485 or standard RJ45 network interface, standard USB interface | ||||||||
Power supply | 380V+10%/50HZ+1,three-phase-five-wire system (3P+N+G) | ||||||||
Meet the testing method | GJB150.5-2009, GB/T 2423-2008; The optional high and low temperature test function also meets the requirements of GB2423.1-2008, GB2423.2-2008, GJB150.3-2009, GJB150.4-2009. |
||||||||
Safety precautions | Power supply undervoltage, overvoltage, phase loss, and phase sequence error protection; Triple over temperature protection; Heater short circuit protection, refrigeration unit overpressure/ overload/ oil shortage protection; humidifier fault protection; Overload/overheating protection for blower motor, overcurrent protection and leakage protection for control board; Cooling water shortage protection, gas source undervoltage protection, etc. | ||||||||
Optional | 1.Optional high and low temperature testing function; 2. Optional special test holes and special shelves; 3. Optional liquid nitrogen function. |